MRT-10 Insulation Resistance Tester
• Covers IR/DAR/PI/DD/SV full-function test modes, and realizes dual-mode calculation of DAR absorption ratio.
•The three-in-one operation matrix of key control, touch screen, and knob mouse simplifies the operation link to the extreme, making the test more flexible and faster.
• Multi-level voltage range can be switched arbitrarily, and custom voltage mode is supported.
• Real-time recording and visualization of insulation resistance-voltage-current time curve, supporting transient analysis of insulation state.
• With AC and DC voltage measurement function, it can automatically identify voltage type and measure AC frequency.
•With cable length measurement function, equipped with controllable hardware and software filters, the anti-interference performance is significantly improved.
• Three safety barriers, including independent discharge circuit, sample breakdown protection and arc detection, not only protect the equipment safety, but also accurately lock the insulation weak points to prevent accidents.
•High and low voltage circuit full isolation architecture to ensure the safety of operators.
• With USB storage function, data can be exported to USB flash drive in batches for easy backup and migration.
|
Test Voltage |
100V, 250V, 500V, 1000V, 2500V, 5000V, 10000V |
|
Full-scale resistance value corresponding to the test voltage |
200GΩ, 500GΩ, 1.00TΩ, 2.00TΩ, 5.00TΩ, 10.0TΩ, 20.0TΩ |
|
Custom test voltage |
50V-10000V, minimum step 10V; when using rotating mouse to control, the step value automatically switches between 10V, 100V, and 200V with the speed, and the higher the speed, the larger the step. |
|
Voltage output accuracy |
0%~+10% |
|
Resistance range and resolution |
0-9.99kΩ 0.01kΩ; 10.0-99.9kΩ 0.1kΩ;
100-999kΩ 1kΩ; 1.00-9.99MΩ 10kΩ; 10.0-99.9MΩ 100kΩ; 100-999MΩ 1MΩ; 1.00-9.99GΩ 10MΩ; 10.0-99.9GΩ 100MΩ; 100-999GΩ 1GΩ; 1.0-9.99TΩ 10GΩ; 10.0-20.0TΩ 100GΩ. |
|
Short circuit current |
1mA-6mA, adjustable, step 1mA
Accuracy: ±5%, continuous output power: 55W. |
|
Test time |
10 seconds to 99 minutes adjustable |
| Voltage measurement | AC/DC 10V-660V, ± (2%+3 digits) |
| AC voltage frequency | 25Hz-85Hz, resolution 0.1Hz, accuracy ±0.2Hz |
| Capacitance measurement |
10nF-25μF, test voltage ≥100V, accuracy: ±10%±5nF |
| Capacitance test
sample discharge rate |
From 10kV to 50V, ≤0.47s/μF |
| AC power supply | 85-264Vrms, 50/60Hz, 65VA |
| Built-in battery | 16.8V, 4400mAH lithium battery |
| Built-in charger | 16.8V, 2.3A |
| Battery life | 10kV@100MΩ load, about 5 hours |
| Use temperature | -10℃~50℃ |
| Relative humidity | <90%, no condensation. |
| Dimensions | 320mm*280mm*200mm |
|
Weight |
About 4.2kg |
| Voltage measurement | AC/DC 10V-660V, ± (2%+3 digits) |
| AC voltage frequency | 25Hz-85Hz, resolution 0.1Hz, accuracy ±0.2Hz |